UNIT 1 – UV–Visible Spectroscopy and Fluorimetry Notes

In pharmaceutical sciences, analytical techniques play a decisive role in identifying, quantifying, and ensuring the quality of drugs. Among these techniques, UV–Visible spectroscopy and fluorimetry stand out for their simplicity, sensitivity, and wide-ranging applications. Unit 1 introduces these two powerful spectroscopic methods, explaining their theoretical foundations, instrumentation, and practical uses in pharmaceutical analysis and research.

Download UNIT 1 – UV–Visible Spectroscopy and Fluorimetry Notes

Get simplified revision notes for this unit:

⬇️

Download Unit 1 Notes PDF

UV–Visible Spectroscopy: Principles and Theoretical Foundations

Electronic Transitions and Absorption of Radiation

UV–Visible spectroscopy is based on the absorption of ultraviolet or visible radiation by molecules, leading to electronic transitions. When a molecule absorbs energy, electrons move from a lower-energy orbital to a higher-energy orbital. Common transitions include σ→σ*, n→σ*, π→π*, and n→π* transitions. Among these, π→π* and n→π* transitions are most relevant in pharmaceutical compounds due to the presence of double bonds and heteroatoms.

Chromophores and Auxochromes

A chromophore is the part of a molecule responsible for absorbing radiation, such as carbon–carbon double bonds or carbonyl groups. Auxochromes, including hydroxyl and amino groups, do not absorb strongly by themselves but modify the absorption characteristics of chromophores. The presence of auxochromes often increases absorption intensity and shifts the wavelength of maximum absorption.

Spectral Shifts and Solvent Effects

Changes in absorption maxima are known as spectral shifts. A shift toward longer wavelengths is called a bathochromic (red) shift, while a shift toward shorter wavelengths is termed a hypsochromic (blue) shift. Solvents significantly influence absorption spectra by stabilizing different electronic states, particularly in polar compounds. Understanding solvent effects is essential when selecting solvents for spectroscopic analysis.

Beer–Lambert Law: Quantitative Basis of UV Spectroscopy

Law and Its Derivation

The Beer–Lambert law states that absorbance is directly proportional to the concentration of the absorbing species and the path length of the sample cell. Mathematically, it is expressed as A = εcl, where A is absorbance, ε is molar absorptivity, c is concentration, and l is path length. This relationship forms the foundation for quantitative drug analysis.

Deviations from Beer–Lambert Law

In practice, deviations may occur due to high solute concentration, chemical interactions, stray light, or instrumental limitations. These deviations can affect accuracy and must be minimized by appropriate sample dilution, proper instrument calibration, and controlled experimental conditions.

Instrumentation of UV–Visible Spectroscopy

Sources of Radiation

Common radiation sources include deuterium lamps for the UV region and tungsten lamps for the visible region. These sources provide stable and continuous radiation over the required wavelength range.

Wavelength Selectors

Monochromators and filters are used to select specific wavelengths. Modern instruments employ diffraction gratings to achieve precise wavelength selection, improving resolution and accuracy.

Sample Cells and Detectors

Sample cells are usually made of quartz for UV measurements and glass for visible measurements. Detection of transmitted radiation is achieved using devices such as phototubes, photomultiplier tubes, photovoltaic cells, and silicon photodiodes. Among these, photomultiplier tubes are highly sensitive and widely used in pharmaceutical instruments.

Applications of UV–Visible Spectroscopy in Pharmaceuticals

Quantitative and Qualitative Analysis

UV–Visible spectroscopy is extensively used for single-component and multi-component drug analysis. It also plays a vital role in spectrophotometric titrations, where changes in absorbance indicate reaction endpoints. These applications make it indispensable for routine quality control.

Fluorimetry: An Advanced and Sensitive Analytical Technique

Theory of Fluorescence

Fluorimetry is based on the emission of light by a substance after absorbing radiation. When electrons return from an excited singlet state to the ground state, fluorescence occurs. Concepts such as singlet, doublet, and triplet electronic states explain the energy transitions involved in fluorescence phenomena.

Internal and External Conversion

Internal conversion refers to non-radiative energy loss within a molecule, while external conversion involves energy loss due to collisions with solvent molecules. These processes influence fluorescence intensity and analytical sensitivity.

Factors Affecting Fluorescence and Quenching

Chemical and Environmental Factors

Fluorescence is affected by molecular structure, solvent polarity, pH, temperature, and the presence of quenching agents. Quenching occurs when fluorescence intensity decreases due to molecular interactions, impurities, or heavy atoms. Understanding these factors is crucial for accurate fluorimetric measurements.

Instrumentation and Applications of Fluorimetry

Fluorimetric Instrumentation

Fluorimeters use intense radiation sources, wavelength selectors, and highly sensitive detectors arranged at right angles to minimize interference from incident light. Proper control of experimental conditions enhances sensitivity and reproducibility.

Applications in Pharmaceutical Analysis

Fluorimetry is particularly valuable for analyzing compounds present in very low concentrations. It is widely used in the estimation of vitamins, antibiotics, steroids, and biological samples, where UV spectroscopy may lack sufficient sensitivity.

Leave a Comment

Your email address will not be published. Required fields are marked *

Scroll to Top