Analytical chemistry plays a central role in pharmaceutical sciences, enabling scientists to identify, quantify, and ensure the quality of drugs and excipients. Unit 2 focuses on instrumental techniques that span molecular structure elucidation, elemental analysis, and measurement of suspended particles. Infrared spectroscopy, flame photometry, atomic absorption spectroscopy, and nepheloturbidometry together provide a powerful toolkit widely used in pharmaceutical analysis, clinical laboratories, and quality control. This article presents a clear, news-style overview of these techniques and their applications.
Download UNIT 2 – Modern Analytical Techniques Notes
Get simplified revision notes for this unit:
⬇️
Download Unit 2 Notes PDF
Infrared (IR) Spectroscopy: Exploring Molecular Vibrations
Introduction to IR Spectroscopy
Infrared spectroscopy is a fundamental analytical technique used to identify functional groups in organic and inorganic compounds. It is based on the absorption of infrared radiation by molecules, which causes changes in their vibrational energy levels. Because each functional group absorbs IR radiation at characteristic frequencies, IR spectra act as molecular “fingerprints.”
Fundamental Vibrational Modes
Polyatomic molecules exhibit two main types of vibrations: stretching and bending. Stretching vibrations involve changes in bond length, while bending vibrations involve changes in bond angle. These vibrations can be symmetric or asymmetric, depending on the movement of atoms within the molecule. The number and type of vibrational modes depend on molecular structure and symmetry.
Sample Handling in IR Spectroscopy
Sample preparation varies with physical state. Solid samples may be analyzed using potassium bromide (KBr) pellets or mull techniques, liquids as thin films between salt plates, and gases in specialized gas cells. Proper sample handling is critical to avoid moisture interference and ensure accurate spectral interpretation.
Factors Affecting Vibrations
Vibrational frequencies are influenced by bond strength, atomic masses, hydrogen bonding, conjugation, and ring strain. Environmental factors such as temperature and solvent interactions can also cause slight shifts in absorption bands.
Instrumentation and Detectors
An IR spectrometer consists of a radiation source, wavelength selector, sample compartment, detector, and recorder. Common detectors include the Golay cell, bolometer, thermocouple, thermistor, and pyroelectric detectors. These detectors convert absorbed IR energy into measurable electrical signals.
Applications of IR Spectroscopy
IR spectroscopy is widely used for drug identification, detection of impurities, compatibility studies, and confirmation of functional groups in raw materials and finished products.
Flame Photometry: Measuring Alkali and Alkaline Earth Metals
Principle of Flame Photometry
Flame photometry, also known as flame emission spectroscopy, is based on the principle that atoms emit light at characteristic wavelengths when excited in a flame. The intensity of emitted light is directly proportional to the concentration of the element in the sample.
Interferences in Flame Photometry
Interferences may arise from other elements, flame temperature variations, or chemical interactions that suppress or enhance emission. Proper calibration and use of internal standards help minimize these effects.
Instrumentation
A flame photometer consists of a nebulizer, flame burner, optical system with filters, detector, and readout device. The flame acts as both atomizer and excitation source.
Applications
Flame photometry is commonly used for the estimation of sodium, potassium, calcium, and lithium in pharmaceutical formulations, biological fluids, and electrolyte solutions.
Atomic Absorption Spectroscopy (AAS): Precision in Elemental Analysis
Principle of Atomic Absorption Spectroscopy
Atomic absorption spectroscopy measures the absorption of radiation by free atoms in the gaseous state. When a sample is atomized, atoms absorb light of specific wavelengths emitted by a hollow cathode lamp corresponding to the element of interest.
Interferences in AAS
AAS can be affected by spectral, chemical, and ionization interferences. Techniques such as background correction and use of releasing agents help improve accuracy.
Instrumentation
The key components of an AAS system include a hollow cathode lamp, atomizer (flame or graphite furnace), monochromator, detector, and signal processor. Graphite furnace AAS offers higher sensitivity for trace analysis.
Applications
AAS is extensively used for detecting trace metals such as iron, zinc, copper, lead, and cadmium in drugs, raw materials, water, and biological samples, making it vital for toxicity and compliance testing.
Nepheloturbidometry: Quantifying Suspended Particles
Principle of Nepheloturbidometry
Nepheloturbidometry measures the intensity of light scattered by suspended particles in a solution. While turbidimetry measures the decrease in transmitted light, nephelometry measures scattered light at an angle, offering greater sensitivity.
Instrumentation
A nepheloturbidimeter consists of a light source, sample cell, detector positioned at an angle to the incident beam, and a signal processor. Precise alignment is essential for accurate measurement.
Applications
This technique is widely used for estimating proteins, polysaccharides, microbial growth, and particulate matter in pharmaceutical suspensions and biological samples.